Patient Monitor User Manual Product Specification
- 281 -
Barometric Pressure
Compensation
Automatic
Data Update Period 1 s
Measurement Range CO
2
: 0 to 25 vol%
N
2
O: 0 to 82 vol%
HAL, ENF, ISO, SEV, DES: 0 to 25 vol%
AwRR: 0 to 150 rpm
Resolution CO
2
: 0.1%
HAL, ENF, ISO, SEV, DES: 0.1%
N
2
O: 1%
AwRR: 1 rpm
☆Accuracy- Standard
Conditions
Gas Range Accuracy
CO
2
0 to 15 vol% ±(0.2 vol% + 2% of reading)
N
2
O 0 to 82 vol% ±(2 vol% + 2% of reading)
HAL
ISO
ENF
0 to 8 vol% ±(0.15 vol% + 5% of reading)
SEV 0 to 10 vol% ±(0.15 vol% + 5% of reading)
DES 0 to 22 vol% ±(0.15 vol% + 5% of reading)
The accuracy specifications above are valid for dry single gases at 22 ± 5 °C and 1013 ± 40
hPa.
☆Accuracy- All
Conditions
Gas Accuracy
CO
2
±(0.3 kPa + 4% of reading)
N
2
O ±(2 kPa + 5% of reading)
Agents ±(0.2 kPa + 10% of reading)
(The accuracy specification for IRMA AX+ is not valid
if more than two agents are present in the gas mixture)
The accuracy specification above is valid for all specified environmental conditions except for
interference specified in Section Effects From Water Vapor Partial Pressure On Gas Readings
and Section Interfering Gas Effects.
☆AwRR Accuracy
±1 rpm
☆Apnea Alarm Delay
20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Work Mode Measure