Patient Monitor User Manual Alarm Information
- 64 -
Message Cause Alarm Level Action Taken
O
2
Cali Required
(Masimo AG module)
O
2
sensor requires
calibration.
Medium
AG Software Error
(Masimo AG module)
Malfunction in the AG
software.
High
AG Hardware Error
(Masimo AG module)
Malfunction in the AG
hardware.
High
AG Uncalibrated
(Masimo AG module)
AG module calibration is
not completed.
High
AG Calibration Fail
(Masimo AG module)
Calibration of the
sidestream AG module fails.
Medium
AG Replace Adapter
(Masimo AG module)
Replacement of the adapter
is required.
Medium Replace the adapter.
AG TEMP Out Of
Range
(Masimo AG module)
The temperature of the AG
module exceeds the
specified working
temperature range.
High
Make sure the AG
module is used
within the specified
temperature range.
Sample Line
Occluded
The sampling tube is
occluded.
Medium
Replace the sampling
tube.
O
2
Sensor Error
(Masimo AG module)
Malfunction in the O
2
sensor inside the sidestream
AG module.
High
Stop measuring
function of AG
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
AG No Adapter
(Masimo AG module)
No adapter is connected. Medium
Connect the adapter
correctly.
No Sample Line
(Masimo AG module)
No sampling tube is
connected.
Medium
Connect the
sampling tube
correctly.
AG Occlusion
AG module sample line
occluded
High
Replace the sampling
line.