Patient Monitor User Manual Alarm Information
- 66 -
Message Cause Alarm Level Action Taken
CO
2
Check Adapter
1) For the Respironics CO
2
module: The cannula is off
or disconnected.
2) For the EDAN CO
2
module: The water trap is
disconnected or not properly
connected.
Low
1) For the
Respironics CO
2
module: Check
whether the adapter
is properly connected
or replace the
adapter.
2) For the EDAN
CO
2
module:
Properly connect the
water trap.
CO
2
Sensor Over
Temp
CO
2
sensor temperature
exceeds +40 °C.
High
Stop using
measuring function
of CO
2
module,
notify biomedical
engineer.
CO
2
Out Of Range
The CO
2
concentration
exceeds the accuracy range
of CO
2
module.
High
Reduce CO
2
concentration.
EtCO
2
Overrange
The EtCO
2
concentration
exceeds the measurement
range.
High
Please check the
monitor or patient
status and adjust the
gas concentration
accordingly.
FiCO
2
Overrange
The FiCO
2
concentration
exceeds the measurement
range.
High
CO
2
Sensor Faulty
CO
2
module failure High
Stop measuring
function of CO
2
module, notify
biomedical engineer.
CO
2
Occlude
Water trap of SideStream is
occluded.
High
Make sure the gas
exhaust works well
CO
2
Replace Adapter
Replacement of the adapter
is required.
Medium
Replace the adapter.