SE-1200 Series Electrocardiograph Service Manual Safety Guidance
- 8 -
WARNING
31. The potential equalization bar can be connected to that of other equipment when
necessary. Make sure that all the equipment is connected to the potential equalization
terminal.
2.2.2 Li-ion Battery Care Warnings
WARNING
1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to
be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It
is necessary to read the user manual carefully and pay more attention to warning
messages.
2. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification as manufacturer configuration should be used.
3. DANGER OF EXPLOSION -- Do not reverse the anode and the cathode when
installing the battery.
4. Do not heat or splash the battery or throw it into fire or water.
5. Do not destroy the battery; Do not pierce battery with a sharp object such as a needle;
Do not hit with a hammer, step on or throw or drop to cause strong shock; Do not
disassemble or modify the battery.
6. When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If
the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean
water first and go to see a doctor immediately.
7. Properly dispose of or recycle the depleted battery according to local regulations.
8. Only when the device is off can the battery be installed or removed.
9. Remove the battery from the electrocardiograph when the electrocardiograph isn’t
used for a long time.
10. If the battery is stored alone and not used for a long time, we recommend that the
battery be charged at least once every 6 months to prevent overdischarge.