SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 38 -
Figure 5-11 Recorder Screen
8. UART Test
On the System Test screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to
move the cursor to the UART item, and then press Enter to open the UART screen.
On this screen, pressing the PRINT/STOP key can make the system send and receive characters.
If the character received is always the same as the character sent, which is displayed on the screen,
it indicates that the systems UART works well.
Press PRINT/STOP again to stop testing. Press ESC to exit the screen.
NOTE: During the testing course, you should not press the Esc key to exit.