SE-1200 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 51 -
3) Disassembly Procedures of the Upper Unit (SE-1200 Express)
Disassemble the upper unit in the following sequence: display screen assembly -> keyboard
assembly -> indicator assembly -> recorder casing -> upper casing.
Remove the four groups of fixing bars and the cross recessed pan head cap screws , and
then the display screen assembly can be dismantled from the upper casing . Remove the two
cross recessed pan head screws , and then the inverted transformer of the display screen can
be dismantled. Remove the fifteen cross recessed pan head self-tapping screws , and then the
keyboard PCB and the silicone keyboard can be dismantled from the upper casing .
Remove the two cross recessed pan head self-tapping screws , and then the indicator board
and the indicator cover can be dismantled from the upper casing . Remove the screws
connecting the upper casing and the recorder casing, and then the recorder casing can be
dismantled from the upper casing .