Note: The above three parameters may be programmed to be
displayed in any desired sequence.
S_HEAT DISPLAY: Cycles on and off to show when
power is applied to Saturator
heater.
BAR GRAPH DISPLAY: The Bar Graph displays a picture of the
dew layer on the chilled mirror surface.
The right hand vertical bar indicates the
mirror itself, and the white bars show the
actual dew layer. The layer can be seen
to vary in thickness as the control loop
brings the mirror into control, and then
maintains a thin layer of dew as the
mirror temperature tracks the dew point
temperature.
BOTTOM ROW:
AIR TEMP. SETPOINT The programmable AT setting
RH SETPOINT The programmable RH setting