EasyManua.ls Logo

Edgetech RH CAL - 15 Specifications

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
15.0 SPECIFICATIONS
RH and AT Operating Ranges
RH Range: 5 % to 95 % for AT = 15 to 50 C
15 % to 95 % for AT = < 15 C
AT Range: 10 °C to 50 °C
RH Accuracy
Range: 5 % to 50 % +/- 0.5 RH %
51 % to 75 % +/- 1.0 RH %
> 75% +/- 1.5 RH %
Dew/Frost Point and Ambient Temperature Range
40 to +60°C (40 to +122°F) D2 Sensor
Dew/Frost Point and Ambient Temperature Accuracy
±0.2°C (±0.36°F) nominal
Dew/Frost Point and Ambient Temperature Sensors
3-wire Platinum Resistance Thermometer (PRT), 100 ohms at 0°C nominal
Precision
0.1 degrees C or F
Chilled Mirror Dew Point Sensor:
Depression
60°C (113°F), nominal, D2 Sensor
Sensor Materials
Chromium, glass, epoxy, anodized aluminum
Slew Rate
1.0°C (1.8°F)/second max., above 0°C
Repeatability
±0.05°C (0.09°F)
Hysteresis
Negligible
Operating Temperature
Control Unit: 0 to +50°C (+32 to +122°F)