© Edwards Limited 2013. All rights reserved. Page 3
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
B756-01-880 Issue K
2Technical data
2.1 General
2.2 Pumping media
Tab le 1 - Gen era l
Performance See Ta b l e 4
Dimensions See Figure 1, 2, 3 and 4
Maximum inlet flange temperature 70 °C
Maximum magnetic field 3.5 mT Horizontal field
Installation category EN61010 part 1, Category 1
Pollution degree EN61010, Pollution degree 2
Electrical supply 24 V d.c.
Equipment type Fixed equipment, for indoor use only
Ambient operating temperature range 5 ºC to 40 ºC
Ambient operating humidity range 10 to 90% RH (non-condensing)
Maximum operating altitude 2000 m
Ambient storage temperature range -30 ºC to 70 ºC
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases.
Do not use Split Flow Pumps to pump explosive gas mixtures as the pumps are not suitable for this
purpose.
If pyrophoric gases are pumped then the customer must supply adequate dilution with an inert gas.
It is the customer’s responsibility to leak check the system to ensure process gases cannot escape
to the atmosphere and cause a hazard.
Gases pumped from different inlets may combine to cause dangerous mixtures.