Fault Warning 56.11 - Exh Temp High
Cause Exhaust temperature high.
Remedy Contact us.
Fault Warning 56.13 - Sensor Missing
Cause Sensor is missing (open circuit).
Remedy Check that the sensor is installed and is correctly connected.
Fault Warning 56.13 - Sensor Fault
Cause Sensor is faulty (short circuit).
Remedy Contact supplier.
Fault Warning 57.11 - DP Mtr Temp High
Cause Pump temperature high.
Remedy Check that cooling water is connected, switched on and is to specicaon giv
en in
Cooling water data on page 21.
Fault Warning 57.13 - Sensor missing
Cause The dry pump end cover temperature sensor has possibly become disconnected
or failed.
Remedy Check that the sensor is installed and is correctly connected.
Fault Warning 57.13 - Sensor Fault
Cause Sensor is faulty (short circuit).
Remedy Contact us.
Fault Warning 63.11 - DP Temp High
Cause The internal temperature of the pump is too high.
Remedy Check that cooling water is connected, switched on and is to specicaon giv
en in
Cooling water data on page 21.
Fault Warning 63.13 - Sensor missing
Cause The dry pump temperature sensor has possibly become disconnected or failed
Remedy Contact us.