10
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
8. RESERVE OXYGEN SYSTEM
As a precaution, EICKEMEYER® may supply you with a reserve oxygen system. If your unit loses electrical power or fails to
operate correctly, the patient Alert System will sound to signal you to switch to your reserve oxygen system (if provided) and
contact EICKEMEYER®.
9. CARING AND PERIODIC MAINTENANCE FOR YOUR UNIT
Note!
Use no lubricants, oils or grease.
Warning!
Before attempting any cleaning procedures, turn the unit off.
Oxygen Humidi er (reusable Bottles)
If your physician has prescribed humidifi cation, clean your humidifi er bottle daily. Follow the instructions supplied by the
manufacturer. If no cleaning instructions were supplied, follow these steps:
1. Wash the humidifi er in a solution of hot water and dishwashing detergent.
2. Soak the humidifi er in a solution of one part white vinegar to three parts hot water for 30 minutes. This solution acts as
a germicidal agent.
3. Rinse thoroughly with warm tap water and refi ll with distilled water for use. Do not overfi ll.
Cannula, Mask and Tubing
Clean and replace the cannula, mask, and tubing as instructed by EICKEMEYER®.
Air Filter and Oxygen Outlet Connector
The connector should not be cleaned at least once a week. To clean, follow these steps:
1. Remove the oxygen outlet connector (if used).
2. Wash in a solution of warm water and dishwashing detergent.
3. Rinse thoroughly with warm tap water and towel dry. The fi lter should be completely dry before reinstalling.
Air lter need to be replaced after device operates each 3,000 hours. Air lter is disposable after used. Not replace the air fi lter
regularly will affect the performance of concentration.
Caution!
To prevent product damage, do not attempt to operate the unit without the air lter or while the lter is still damp.