USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
10. MINOR TROUBLESHOOTING
Symptom Possible Cause Remedy
Unit does not operate. Power light is off
when the power switch is on”. Audible
alarm is pulsing and Service Required
light is fl ashing.
Power cord not properly inserted into
wall outlet.
Check power connection at the wall
outlet. On 230 volt units, check the back
of the unit.
No power at wall outlet. Check your home circuit breaker and
reset if necessary. Use a different wall
outlet if the situation occurs again.
Oxygen concentrator circuit breaker
activated (selected units).
Press the concentrator circuit breaker
button (if equipped) located below the
power switch. Use a different wall
outlet if the situation occurs again.
Unit operates; the Power light is on
when the Power switch is On”. Red
Service Required light is illuminated.
Audible alarm may be sounding.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area; make sure
there is nothing restricting the unit
exhaust.
Blocked or defective cannula, catheter,
face mask or oxygen tubing.
Detach cannula, catheter, or face mask.
If proper fl ow is restored, clean or
replace if necessary. Disconnect the
oxygen tubing at the oxygen outlet. If
proper fl ow is restored, check oxygen
tubing for obstructions or kinks.
Replace if necessary.
Blocked or defective humidifi er bottle. Detach the humidifi er from the oxygen
outlet. If proper fl ow is obtained, clean
or replace humidifi er.
Flow meter set too low. Set fl ow meter to prescribed fl ow rate.
If the above remedies do not work,
contact EICKEMEYER®.
Unit operates the power light is on
when power switch is on“, audible
low-frequency vibration sound is
detected.
Turn your unit off”. Switch to your
reserve oxygen system, and contact
EICKEMEYER® immediately.
If any other problems occur with your
oxygen concentrator.
Turn your unit Off. Switch to your
reserve oxygen system, and contact
EICKEMEYER® immediately.
Both the green Normal Oxygen and the
yellow Low Oxygen lights are either on
or off.
O.C.I malfunction Contact EICKEMEYER®.
Yellow Low Oxygen light is on or the
yellow low Oxygen light is on and the
intermittent audible Signal is sounding.
Flow meter is not properly yet. Ensure the fl ow meter is properly set to
the prescribed.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area: make sure
there is nothing restricting the unit
exhaust.