USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
6. BEFORE OPERATING YOUR OXYGEN CONCENTRATOR
1. Before operating your unit always check to be sure the air lter (located on the back of your unit) is existed. Proper replace
of this fi lter is discussed in the Caring for Your Concentrator section on page 10.
2. Attach the appropriate oxygen accessories to the oxygen outlet.
Oxygen Tubing Connection:
a. Attach the oxygen tubing directly to the connector (Figure 3).
Figure 3
Oxygen Tubing Connection with Humidifi cation:
If your physician has prescribed an oxygen humidifi er as part of therapy, follow these steps:
a. Fill the humidifi er bottle with distilled water. Do not overfi ll.
b. Attach the oxygen tubing directly to the humidifi er bottle outlet fi tting (Figure 4).
Figure 4
Your physician has prescribed either a nasal cannula, catheter, or face mask, in most cases, they are already attached to the
oxygen tubing. If not, follow the manufactures instructions for attachment.
Remove the power cord completely from the line cord strap. Make sure the power switch is in the “Off” position, and insert the
plug into the wall outlet. The unit is double insulated to guard against electric shock.
Warning!
Improper use of the power cord and plugs can cause a burn, re or other electric shock hazards. Do not use the unit
if the power cord is damaged.