13
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Symptom Possible Cause Remedy
Red Service Required light is on and an
intermittent audible signal is sounding.
Flow meter is not properly yet. Ensure the fl ow meter is properly set to
the prescribed.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area: make sure
there is nothing restricting the unit
exhaust. If the above remedies do not
work, contact EICKEMEYER®.
11. SPECIFICATIONS
No. Item Rating
1. Delivery Rate 0.5 to 5 LPM
2. Outlet Pressure 8.5 – 13 psi (68.5 – 90 kpa)
3. Electrical Rating 230 V, 50 Hz, 1.4A (Max.)
4. Oxygen Percentage 0.5 ~ 5 LPM 93% ± 3%
5. Average oxygen concentration At 2 l/min: 93% ± 3%
6. Operating Environment Range Temperature 10 °C to 40 °C
Humidity 15% to 95%
Atmospheric pressure 50 ~ 106 kpa (7.3 ~ 15.4 psi)
7. Weight 15.3kg
8. Sound Level (ISO 8359:1996 from front) ≤55dbA
9. Dimensions (in mm) L 342 x W 368 x H 572
10. Operating System Time Cycle / Pressure Swing
11. Storage Conditions Temperature -40 °C to 70 °C
Humidity 10% to 100%
Atmospheric pressure 50 ~ 106 kpa (7.3 ~ 15.4 psi)
12. Equipment Class and Type The type of protect against electric shock: Class II
The degree of protect against electric shock: TYPE B Type B Applied part
13. Pressure Relief Valve 276 kPa ± 34.5 kPa
14. Warranty Period limit 1 Year