11
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Exterior Cabinet
Warning!
To avoid electric shock, do not remove the concentrator cabinet. The cabinet should only be removed by a qualifi ed
EICKEMEYER® technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or
cleaning agents.
Clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild house-hold cleaner and wipe it dry.
Periodic Maintenance for your Unit
1. Clean the air fi lter at least once a week. The steps of cleaning see the Air Filter and Oxygen Outlet Connector.
2. Change the felt of the Inlet Pre-Filter at least once half a year. Follow those steps:
Open the Important Safeguard pull out the Inlet Pre-Filter. Pull out the old felts from the Inlet Pre-Filter (two sides). Then
insert the new felts to the Inlet Pre-Filter each side, insert the Inlet Pre-Filter and close the Important Safeguard.
3. Change the Bacterial Filter at least once a year. You can not change this by yourself, you must contact EICKEMEYER®
change this for you.
Note!
• The use of certain humidifi ers and accessories may impair the oxygen concentrator’s performance.
• If the Service Required light illuminates and the audible signal alarms but the unit is not operating, there is no
power to the unit. Refer to the Minor Troubleshooting chart on page 12 and contact EICKEMEYER®.