27 | Avoset Quick Guide
List of Alarms
ALARM TEXT CAUSE TROUBLESHOOTING ALARM PRIORITY
INFUSION PROGRESS
INFUSION COMPLETE
The VTBI has been delivered Replace the bag or turn off the pump
(see page 20)
High-priority Alarm
CHANGE BAG NOW
The VTBI has been delivered Replace bag High-priority Alarm
ACTION INCOMPLETE
The infusion has not started yet, or an
action issued is incomplete, and no
key was pressed for a predened time
duration
Start the infusion or address the incomplete
action
Low-priority Alarm
PUMP IS PAUSED
The infusion has already started and
the pump was paused for 30 seconds
Press the Start/Pause button if the infusion
should be resumed (see page 30)
Low-priority Alarm
TIME TO CHANGE BAG
Infusion bag is almost empty Replace bag Low-priority Alarm
BATTERY
BATTERY DEPLETED
Battery power is too low to operate the
pump
Replace batteries (see page 34) High-priority Alarm
(Screen turns off with
a blinking Red
light and a beep)
Batteries were removed or the pump
suddenly lost power during operation
Replace batteries (see page 35).
Verify batteries are installed correctly
High-priority Alarm
LOW BATTERY
The batteries are low Replace batteries (see page 33) Low-priority Alarm