28 | Avoset Quick Guide
List of Alarms
ALARM TEXT CAUSE TROUBLESHOOTING ALARM PRIORITY
PUMP MALFUNCTIONS
ERROR 1 REPLACE PUMP
A critical internal error Do not use pump.
Patients: contact your service provider, pump
should be replaced
Non patients: turn off pump and send it for
service
High-priority Error
ERROR 2 REPLACE PUMP
A non-critical internal error Do not use pump.
Patients: contact your service provider, pump
should be replaced
Non patients: send a new infusion program to
the pump
High-priority Error
ERROR 3 REPLACE PUMP
An internal clock error Do not use pump.
Contact your service provider
High-priority Error
ERROR 4 REPLACE PUMP
Pump software update failure Do not use pump.
Patients: contact your service provider, pump
should be replaced
At service: update pump software
High-priority Error
RESTART PUMP
An internal error preventing pump
operation
Restart pump. If the alarm occurs for the second
time, replace the pump and contact your
service provider
High-priority Alarm
HIGH TEMPERATURE
Pump overheating Turn off pump and let it cool down, if the
temperature is still too high when the pump is
turned on, the alarm will occur again
High-priority Alarm
RECOVERED FROM ERROR
Pump restarted after an internal error The interrupted infusion cannot be resumed.
VTBI was reset to the initially programmed value.
If you start the infusion, it will not resume from
where it stopped before interrupted
High-priority Alarm