35 | Avoset Quick Guide
Battery Extracted
A sudden power depletion or battery disconnection while the pump is on.
+
–
START
BOLUS
PRIME
PAUSE
ON/OFF
GO
RUN
ALARM
1. The pump is turned off
Alarm light
is Red
2. Press down the
battery cover latch to
open it
3. Remove each
battery starting with
the positive (+) ends
5. Close the battery cover.
To resume infusion, turn
the pump on and select
RESUME multi-function
button (see page 19).
4. Insert 3 AA (1.5V) batteries according the diagram
in the battery compartment.
Slide the negative (-) end rst, compress the spring
and snap the positive (+) end into place
BEEP
Infusion is saved when batteries are
removed, and can be resumed later.