34 | Avoset Quick Guide
+
START
BOLUS
PRIME
PAUSE
ON/OFF
GO
RUN
ALARM
+
START
BOLUS
PRIME
PAUSE
ON/OFF
GO
RUN
ALARM
Battery Depleted
2. The pump is turned off1. Press the
SHUTDOWN
multi-function button
BEEP
After 3 minutes
without a proactive
turn off of the pump,
it will automatically
turn off
The batteries are about to deplet in 3 minutes.
3. Press down the
battery cover latch to
open it
4. Remove each
battery starting with
the positive (+) ends
6. Close the battery cover.
To resume infusion, turn the
pump on and select
RESUME
multi-function button (see
page 19).
5. Insert 3 AA (1.5V) batteries according the diagram
in the battery compartment.
Slide the negative (-) end rst, compress the spring
and snap the positive (+) end into place
OO:O3
BATTERY
DEPLETED
SHUTDOWNMUTE
HOURS MINUTES
Infusion is saved when batteries are
removed, and can be resumed later.