MAINTENANCE
DEVICE MAINTENANCE
Warning!
The operator shall ensure completion of repeated testing of the device at least once every 24
months (EN 62353) or at intervals defined by applicable national legal regulations. A record
of these test results shall be completed (e.g. per EN 62353, Annex G) together with the
measurement methods.
The device has been designed and manufactured to keep maintenance to a minimum. The following
work must be performed to retain the proper and reliable operation of the compressor.
Before starting compressor maintenance work, it is necessary to check it the
compressor can be disconnected from the device to ensure that the person using the
device is not at risk of health damage and there is no risk of any other material
damages!
Air pump components (head, cylinder, pressure hose, etc.) are very hot during and
shortly after compressor operation do not touch these components!
Repair works beyond normal maintenance can be performed only by qualified
personnel or he manufacturer’s customer service.
Use only spareparts and accessories approved by the manufacturer.
PROTECT EYESIGHT, WEAR GOGGLES, WHEN VENTING COMPRESSED AIR FROM
THE COMPRESSED AIR CIRCUIT (AIR TANK).
The 8-10 bar compressor model is equipped with an hour counter (Fig. 3).
The work below may only be performed by trained personnel as follows:
BEFORE STARTING ANY MAINTENANCE WORK, FIRST DISCONNECT THE
COMPRESSOR FROM THE MAINS (PULL OUT THE MAINS PLUG) AND VENT THE
COMPRESSED AIR FROM THE AIR TANK.
Allow the equipment to cool down before maintenance, service or
connecting/disconnecting the compressed air supply!
Reconnect the grounding lead if removed during service work to its previous location
once the work is complete.
For any maintenance or repair work, the compressor may be removed from the cabinet by pulling it
from the cabinet (on rolling casters as much as the cord between the compressor and the cabinet
allows). See also chap. 19.). The required maintenance or repair work may then be performed.
If the cord between the compressor and the cabinet becomes disconnected (by pulling the mains
plug, chap. 14.2.), the compressor shall not be functional. In order to bring the compressor into
function again, please follow chap. 14.2. or 19.10. or 19.11.

Table of Contents

Other manuals for EKOM DK50-10 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EKOM DK50-10 S and is the answer not in the manual?

EKOM DK50-10 S Specifications

General IconGeneral
BrandEKOM
ModelDK50-10 S
CategoryAir Compressor
LanguageEnglish

Related product manuals