13
5. Safety notes
Please observe the following instructions under all circumstances
> Before using the Elcomed for the first time, store it at room temperature for 24 hours.
> Check the Elcomed, the straight or contra-angle handpiece and the motor with cable for damage and loose parts each time
before using. Correct any faults or refer to an authorized W&H service partner (see page 72). Do not operate the Elcomed
if it is damaged.
> Check the parameter settings every time the device is restarted.
> Perform a test run each time before using.
> Make sure that the operation can be completed safely even if a device or instrument malfunctions.
> Never touch the chuck mechanism of the straight or contra-angle handpiece during operation or while they are still
running down.
> Only fit the straight or contra-angle handpiece onto the motor, when the motor is at a complete standstill.
> Never touch the rotary instrument which is still rotating.
> Always ensure correct operating conditions and that sufficient and adequate cooling is delivered.
> Avoid overheating at the treatment site.
> When changing the fuse, disconnect the device from the power supply and only use W&H original fuses.
> Never touch the patient and the connection for the foot control simultaneously.
> The ESD spring contact on the underside of the foot control must touch the floor during operation.
> Moisture in the plug or motor can lead to a malfunction. (Risk of short circuit)
Use only suitable and serviceable tools
Ensure that you comply with the manufacturers instructions for surgical straight and contra-angle handpieces with respect
to maximum speed, maximum torque, clockwise and anticlockwise rotation.
Inappropriate use
Improper use, in addition to incorrect assembly, installation, modification or repairs of the Elcomed or failure to comply with
our instructions invalidates all claims under warranty and any other claims.