Hygiene and maintenance
Automated cleaning and disinfection
Motor with cable / Motor support / Stand
W&H recommends automated cleaning and disinfection using a washer-disinfector (WD).
Read the notes, follow the instructions and heed the warnings provided by the manufacturers of washer-disinfectors,
cleaning agents and/or disinfectants.
> The control unit and foot control are not approved for automated cleaning and disinfection.
Evidence of the basic suitability of the motor with cable, the motor support and the stand for effective automated
disinfection was provided by an independent test laboratory using the »Miele PG 8582 CD« washer-disinfector (Miele &
Cie. KG, Gütersloh) and the »Dr. Weigert neodisher® MediClean forte« cleaning agent (Dr. Weigert GmbH & Co. KG, Hamburg).
> Cleaning at 55°C (131°F) – 5 minutes
> Disinfection at 93°C (200°F) – 5 minutes