56
Motor with cable / Motor support
W&H recommends sterilization according to EN 13060, EN 285 or ANSI/AAMI ST79.
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of steam sterilizers.
> The program selected must be suitable for the motor with cable and the motor support.
Recommended sterilization cycles
> Steam sterilization (type B, S, N)
> Sterilization time at least 3 minutes at 134°C (273°F), 30 minutes at 121°C (250°F)
> Maximum sterilization temperature 135°C (275°F)
Evidence of the motor with cable and the motor support basic suitability for effective sterilization was provided by an
independent test laboratory using the LISA 517 B17L steam sterilizer (W&H Sterilization S.r.l., Brusaporto (BG)), the Systec
VE-150 steam sterilizer (Systec) and the CertoClav MultiControl MC2-S09S273 steam sterilizer (CertoClav GmbH, Traun).
“Dynamic-air-removal prevacuum cycle” (type B): temperature 134°C (273°F) – 3 minutes*
“Steam-flush pressure-pulse cycle” (type S): temperature 134°C (273°F) – 3 minutes*
“Gravity-displacement cycle” (type N): temperature 121°C (250°F) – 30 minutes**
* according to EN 13060, EN 285, ISO 17665
** according to ANSI/AAMI ST55, ANSI/AAMI ST79
Hygiene and maintenance
Sterilization