5 GETTING STARTED (continued)
Option
Action Required
ABOUT
Press SEL to view the ABOUT menu.
RESET
Press SEL to view the RESET menu
AUTO SWITCH
OFF
Press SEL followed by + or - to set the auto switch off delay; 1
to 15 minutes or off (X). Press OK when nished.
OPENING
SCREEN
Press SEL to toggle the opening screen on or off.
VOLTAGE
LOCKED
Press SEL to toggle the voltage lock on or off, see Section 5.6
‘Voltage and Sensitivity Locks’ on page en-11.
SENSITIVITY
LOCKED
Press SEL to toggle the sensitivity (current) lock on or off, see
Section 5.6 ‘Voltage and Sensitivity Locks’ on page en-11.
www.elcometer.com
en-10
R
5.5 CLICKS, BEEPS, ALARMS AND LIGHTS
The Elcometer 266 emits a range of sounds and lights whilst in use,
see Table 2 below.
Sound
Lights
Indicates
Single beep -
high pitch
Red light on high voltage
probe handle illuminates
High voltage to probe is
switched on
Double beep -
high pitch
Red light on high voltage
probe handle ashes on/off
The safety interlock on the
high voltage probe handle
is not being gripped by your
hand
Clicks - continuous
series of
Red light on high voltage
probe handle is illuminated
High voltage is present at
the probe
Alarm buzzing
Blue light on high voltage
probe handle ashes on/off
Flaw detected