EasyManua.ls Logo

ELGA PC220DIBPM3 - Purified Water Specifications; Feed Water Specifications

Default Icon
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connections
Inlet 8mm (5/16) OD tube
Drain 8mm (5/16) OD tube
Outlet 8mm (5/16) OD tube
Puried Water Specication
Model
PURELAB®
Chorus 2
PC210DIXXM3
PURELAB®
Chorus 2
PC210DIBPM3
PURELAB®
Chorus 2
PC220DIXXM3
PURELAB®
Chorus 2
PC220DIBPM3
Make-up rate @15˚C
(*4bar feed)
10l/h 10l/h 20l/h 20l/h
Delivery ow rate -
maximum
240l/24h 240l/24h 480l/24h 480l/24h
Inorganic Rejection 1 to >10MΩ.
cm
1 to >10MΩ.
cm
1 to >10MΩ.
cm
1 to >10MΩ.
cm
Organics (TOC) - Typical <30ppb <30ppb <30ppb <30ppb
Organics (MW >200 Dalton) >99%
Rejection
>99%
Rejection
>99%
Rejection
>99%
Rejection
Bacteria - Typical <5 CFU/ml <5 CFU/ml <5 CFU/ml <5 CFU/ml
Particles >99%
Rejection
>99%
Rejection
>99%
Rejection
>99%
Rejection
Source Quality Potable mains
supply
Potable mains
supply
Potable mains
supply
Potable mains
supply
Fouling Index-Maximum 10 10 10 10
TDS-Maximum 1000ppm 1000ppm 1000ppm 1000ppm
Conductivity <2000µS/cm <2000µS/cm <2000µS/cm <2000µS/cm
Free/total Chlorine-Maximum 0.5ppm 0.5ppm 0.5ppm 0.5ppm
Chloramine - Maximum 0.5ppm 0.5ppm 0.5ppm 0.5ppm
Heavy Metals-Maximum
(Fe/Mn)
0.05ppm 0.05ppm 0.05ppm 0.05ppm
Silica-Maximum 30ppm 30ppm 30ppm 30ppm
Temperature 1-35˚C 1-35˚C 1-35˚C 1-35˚C
Flowrate (max requirement
@ 15˚C)
100l/h 100l/h 100l/h 100l/h
Drain requirements 80l/h 80l/h 80l/h 80l/h
Maximum feed pressure 6.0 bar
(90 psi)
2.0bar
(30psi)
6.0 bar
(90 psi)
2.0bar
(30psi)
Minimum feed pressure
4.0 bar
(58 psi)
0.5bar
(7.5psi)
4.0 bar
(58 psi)
0.5bar
(7.5psi)
Feed Water Specication
Chorus 2 Operator Manual MANU40003 Version 03 3/19 Page 12
7. TECHNICAL SPECIFICATIONS