4.2 Testing wrist straps separately
DIP switch 1 + 2 at the rear of the unit is set to the "OR" function (default
setting) or the function "only wrist strap".
Put on the wrist strap and connect it via the coil cord to the snap or the
plug (5d, fig. 2) on the left hand side of the unit.
Press the left measuring plate and keep it pressed. An audible signal indi-
cates that the tester is switched on. After a short measuring time the test
result will be indicated by a LED:
The tester will be switched off when the measuring plate is released.
4.3 Testing coil cords separately
A coil cord without a wrist band can be checked as displayed on the front
panel. The connection is made via the 3 mm snap (5c, fig. 2) and the snap
or the plug (5d, fig. 2) on the left hand side of the unit.
DIP switch 1 + 2 at the rear of the unit is set to the "OR" function (default
setting) or the function "only wrist strap".
Press the left measuring plate and keep it pressed. An audible signal indi-
cates that the tester is switched on. After a short measuring time the test
result will be indicated by a LED:
The tester will be switched off when the measuring plate is released.
Green LED "OK" flashes: The resistance is in the range of
750 kOhm to 35 MOhm or 0 to 35 MOhm with
the lower limit switched off.
Red LED "Hi-Fail" flashes:
Audible signal
The resistance exceeds the upper limit
of 35 MOhm.
Red LED "Lo-Fail" flashes:
Audible signal
The resistance is below the lower limit
of 750 kOhm
(
not applicable if the lower limit is switched off).
Green LED "OK" flashes: The resistance is in the range of
750 kOhm to 35 MOhm or 0 to 35 MOhm with
the lower limit switched off.
Red LED "Hi-Fail" flashes:
Audible signal
The resistance exceeds the upper limit
of 35 MOhm.
Red LED "Lo-Fail" flashes:
Audible signal
The resistance is below the lower limit
of 750 kOhm
(
not applicable if the lower limit is switched off).