DIGITAL CONDUCTIVITY METER
The information contained in the accompanying document is proprietary and confidential, and
may not be copied in any manner whatsoever without prior written consent of Emcee
Electronics, Inc. The document and the material therein may not be used for any purpose other
than that intended by Emcee Electronics, Inc.
COPYRIGHT 2005 EMCEE ELECTRONICS, INC.
REVISION DATE: October 23, 2014
Page 8 of 11
8.0 TEST PROCEDURE
CAUTION: This device must be grounded before and after introduction of the probe
into the tank and remain grounded until after complete withdrawal from the tank.
8.1 To initiate a test. depress the pressure sensitive switch once and then again when
EMCEE is displayed. The following will occur:
8.1.1 The red LED located just above the pressure sensitive switch (MC Logo) will
begin to blink indicating that the meter is in the ready mode to perform a test.
8.2 Insert the probe into the sample up to and covering the four holes in the probe.
8.2.1 “READ NEW DATA” will be displayed.
8.2.2 The red LED will stop blinking and remain on indicating that a test is in
progress. Hold the meter steady during the test period.
8.2.3 A broken line (----------) will blink on and off three times (3-seconds) during
the test period.
8.2.4 At the end of the test period, the red LED will extinguish indicating that the test
is complete and the measurements are stored in memory.
8.2.5 “NEW DATA PS/M” will be displayed
8.2.5.1 Value of measurement will be displayed – Example 325
8.2.5.2 The temperature of the sample will be displayed as follows:
8.2.5.2.1 Temperature in Degrees Centigrade - Example: 28.8C
8.2.5.2.2 Temperature in Degrees Fahrenheit - Example: 83.8F
Note: The new data (pS/m and temperature) will replace the previous data and will
be stored in memory until another test is performed.
It is recommended that the probe be rinsed with alcohol followed by toluene prior to
performing another test or storing the meter.