DIGITAL CONDUCTIVITY METER
The information contained in the accompanying document is proprietary and confidential, and
may not be copied in any manner whatsoever without prior written consent of Emcee
Electronics, Inc. The document and the material therein may not be used for any purpose other
than that intended by Emcee Electronics, Inc.
COPYRIGHT 2005 EMCEE ELECTRONICS, INC.
REVISION DATE: October 23, 2014
Page 11 of 11
11. PRECISION*
The electrical parameters of the meter are factory calibrated to 1% of reading.
However, due to variations in product measurement characteristics, the repeatability
(r) and reproducibility (R) are as follows:
PS/M r R
1 1 1
15 6 8
20 7 9
30 9 11
50 13 16
70 15 19
100 19 24
200 29 36
300 37 46
500 51 62
700 62 76
1000 77 94
1500 98 120
PS/M – picosiemens per meter
r = Repeatability - Maximum allowable difference between 2-successive
measurements performed by the same operator using the same meter to test the same
sample.
R = Reproducibility – Maximum allowable difference between 2-measurements
performed on the same sample by 2different operators using 2-different meters.
*Precision is based on a Inter-laboratory Test (ILT) Program conducted in November
2004 under the auspices of the American Society of Test and Materials (ASTM)
International. Since the Program was conducted at room temperature (23ºC – 73.4ºF)
the precision may vary at other temperatures.