3
CROSBY J SERIES PRESSURE RELIEF VALVE WITH FISHER™ 4320 WIRELESS POSITION MONITOR
INSTALLATION, OPERATION AND MAINTENANCE MANUAL
2.2 Calibration
Mounting the Fisher 4320 Wireless 
Position Monitor
1.  Install a battery into the Fisher 4320 
Wireless Position Monitor.
WARNING
Due to the combustible nature of the lithium 
content, the power module has special 
installation, operation, storage, and/or shipping 
requirements. Observe all warnings included with 
the power module before installing, operating, 
storing, or shipping the 4320 Position monitor.
Failure to do so may result in personal injury or 
property damage from fire or explosion.
CAUTION
When installing components, proper means of 
electrostatic discharge protection is required.
Failure to use a grounding strap, or other means 
of electrostatic discharge protection can result in 
damage to the electronics.
CAUTION
To avoid static discharge do not rub or clean the 
antenna with solvents.
Failure to do so may result in personal injury or 
property damage from fire or explosion.
2.  Mount the 4320 to the bracket and lightly 
tighten screws enough to hold in place. The 
screws will be tightened once calibration 
step is complete.
3.  Mount the calibration plate and bracket on 
the pressure relief valve bonnet as shown 
on Figure 2-2.
WARNING
Due to the combustible nature of the lithium 
content, the power module has special 
installation, operation, storage, and/or shipping 
requirements. Observe all warnings included with 
the power module before installing, operating, 
storing, or shipping the 4320 Position monitor.
Failure to do so may result in personal injury or 
property damage from fire or explosion.
4.  Slide the Polytetrafluoroethylene (PTFE) 
bushing sleeve on the magnet bracket. 
The PTFE open slot should face in down 
direction.
5.   Secure the Magnet in place under the bend 
in the Magnet Bracket with 2 screws and 
lock washers provided with Magnet. Torque 
force listed in Table 2, Location 1.
6.  Position the magnet bracket assembly to 
the valve stem at the dimension “A” shown 
in Table 1. Lock in place.
7.  Verify dimension “A” by caliper or depth 
micrometer to ensure it is in the tolerance 
as shown in the Table 1.
8.  Loosen the Screws on 4320 Bracket and line 
up the mark on the back of 4320 with the top 
mark on the magnet. Lock 4320 in place on 
Bracket with quantity of 4, M6 x 10mm HEX 
CAP Screws and Lock Washers with Torque 
force listed in Table 2, Location 2. Center 
the magnet assembly inside the 4320 slot by 
slightly adjusting the magnet bracket, if it is 
needed. This represents the “0% position”, 
valve closed.
CAUTION
Do not install a magnet assembly that is shorter 
than the physical travel of the valve. Loss of 
control will result from the magnet assembly 
moving outside the range of the index mark in the 
feedback slot of the 4320 housing.
Failure to do so may result in personal injury or 
property damage.
FIGURE 2-1
4320 and Bracket Assembly
FIGURE 2-2
Magnet Alignment
FIGURE 2-3
Measurement for Dim A
FIGURE 2-4
Measurement for Dim A