Alert data structure
Transmitter action if condition occurs
Contents Clearing
Alert Statistics One record for each alert (by alert number)
that has occurred since the last master reset.
Each record contains:
• A count of the number of occurrences
• Timestamps for the most recent posting
and clearing
Not cleared; maintained across transmitter
power cycles
Recent Alerts 50 most recent alert postings or alert clearings Not cleared; maintained across transmitter
power cycles
8.5 Read totalizer and inventory values
ProLink III View the desired variable on the main screen under Process Variables.
Field Communicator Service Tools > Variables > Totalizer Control
Overview
Totalizers keep track of the total amount of mass or volume measured by the transmitter
since the last totalizer reset. Inventories keep track of the total amount of mass or volume
measured by the transmitter since the last inventory reset.
Tip
You can use the inventories to keep a running total of mass or volume across multiple totalizer
resets.
8.6
Start and stop totalizers and inventories
ProLink III Device Tools > Totalizer Control > Totalizer and Inventories > Start All Totals
Device Tools > Totalizer Control > Totalizer and Inventories > Stop All Totals
Field Communicator Service Tools > Variables > Totalizer Control > All Totalizers > Start Totalizers
Service Tools > Variables > Totalizer Control > All Totalizers > Stop Totalizers
Overview
When you start a totalizer, it tracks process measurement. In a typical application, its value
increases with flow. When you stop a totalizer, it stops tracking process measurement and
its value does not change with flow. Inventories are started and stopped automatically,
when totalizers are started and stopped.
Transmitter operation
Configuration and Use Manual 99