Emerson MICRO MOTION 1500 - Perform a D1 and D2 Density Calibration Using Prolink III

Emerson MICRO MOTION 1500
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If LD Optimization is enabled on your meter, disable it. To do this, choose Configure
> Manual Setup > Measurements > LD Optimization. LD Optimization is used only
with large sensors in hydrocarbon applications. In some installations, only customer
service has access to this parameter. If this is the case, contact customer service
before continuing.
The calibrations must be performed without interruption, in the order shown. Make
sure that you are prepared to complete the process without interruption.
Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
Restriction
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be
performed on water.
9.7.1 Perform a D1 and D2 density calibration using ProLink III
1. Read the Prerequisites on page 114 if you have not already done so.
2. See the following figure.