Option Description
OUT
x ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is
out of range, that is, less than Setpoint A or greater than Setpoint B, end-
points included.
3. Assign a process variable to the event.
4. Set values for the required setpoints.
For HI and LO events, set Setpoint A.
For IN and OUT events, set Setpoint A and Setpoint B.
5. (Optional) Configure a Discrete Output to switch states in response to the event
status.
6. (Optional) Specify the action or actions that the transmitter will perform when the
event occurs.
With ProLink III: Device Tools > Configuration > I/O > Action Assignment
With the Field Communicator: Configure > Alert Setup > Discrete Events > Assign
Discrete Action
Options for Enhanced Event Action
Action
Label
ProLink III Field Communicator
Standard
None (default)
None None
Start sensor zero
Start Sensor Zero Perform auto zero
Start/stop all totalizers
Start/Stop All Totalization Start/stop totals
Reset mass total
Reset Mass Total Reset mass total
Reset volume total
Reset Volume Total Reset volume total
Reset gas standard volume total
Reset Gas Std Volume Total Reset gas standard volume total
Reset all totals
Reset All Totals Reset totals
Meter verification
Start meter verification test
Start Meter Verification
Not available
Note
Before assigning actions to an enhanced event, check the status of the event. If it is ON, all assigned
actions will be performed when the new configuration is implemented. If this is not accepatable,
wait until an appropriate time to assign actions to the event.
Integrate the meter with the control system
Configuration and Use Manual 79