14 Micro Motion
®
Model 1500 Transmitters with the Filling and Dosing Application
Flowmeter Startup
Figure 3-3 ProLink II – Flowmeter zero procedure
Figure 3-4 Zero button – Flowmeter zero procedure
Modify zero time
if required
Calibration
Failure LED
Calibration in Progress
LED turns red
Green
Troubleshoot
Red
Perform Auto Zero
Done
ProLink >
Calibration >
Zero Calibration
Wait until Calibration in
Progress LED turns green
Status LED
Status LED flashes
yellow
Done
Solid Green or
Solid Yellow
Troubleshoot
Solid
Red
Press ZERO button