Installation Manual
20002298, Rev CH
April 2020
Micro Motion
F-Series Coriolis Flow and
Density Sensors
Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion F300 and is the answer not in the manual?

Emerson Micro Motion F300 Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion F300
CategoryAccessories
LanguageEnglish

Summary

Before you begin

About this document

Provides information on planning, mounting, wiring, and grounding the F-Series sensor.

Hazard messages

Defines hazard message criteria based on ANSI standards Z535.6-2011 (R2017).

Related documentation

Lists additional product documentation available on DVD or online at emerson.com.

Planning

Installation checklist

Ensures hazardous area suitability and verifies temperature limits before installation.

Best practices

Guidance for optimal sensor performance, including pipe run requirements and orientation.

Temperature limits

Details process and ambient temperature ranges for sensor operation and electronics.

Mounting

Mount the sensor

Guidance on physically installing the sensor to minimize torque and bending load on connections.

Attach extended electronics

Instructions for connecting separate electronic components to the sensor case using an extender.

Wiring

Options for wiring

Outlines different wiring procedures based on the installed electronics option.

Connect 4-wire cable

Detailed procedures for preparing and connecting the 4-wire cable to the core processor terminals.

Connect the 9-wire cable

Steps for preparing and connecting the 9-wire cable to the junction box terminals.

Grounding

Prerequisites

Lists guides and requirements for proper grounding practices based on applicable standards.

Procedure

Steps to ensure the meter is properly grounded to earth or the facility's ground network.

Supplementary information

Purge the sensor case

Procedure for purging the sensor case with inert gas (argon or nitrogen) for maintenance.

Pressure relief

Information on rupture disks that vent process fluid in case of a tube breach.

Related product manuals