108 HART Communications
Section 10: HART Communication Instruction Manual
April 2017 LIQ-MAN-1066
Parameters Common to All 1066 Transmitters
APPENDIX 10.3 – 1066 HART Configuration Parameters
Type Enumeration / Range
Read/Write
Default
input_filter_type
Selects between continuous and adaptive filtering of
the measurement.
ENUM Adaptive
Continuous
R / W Adaptive
pv_damping_value
Provides the filter time constant.
FLOAT 0.0 -- 99.0 seconds R / W 0
instrument_software_version
The software version of the main processor.
FLOAT R
input_software_version
The software version of the input processor.
FLOAT R
Display Parameters:
loi_language
Selects the language to be displayed by the 1066.
ENUM English Italiano
Français Português
Español Chinese
Deutsch Russian
R / W English
loi_warnings
Enables or disables the display of transmitter warnings.
ENUM Enable
Disable
R / W Enable
loi_configuration_code
Locks and unlocks access to configuration from the
keypad display of the 1066. "000" disables.
UINT (2) R / W 0
loi_calibration_code
Locks and unlocks access to calibration from the key-
pad display of the 1066. "000" disables.
UINT (2) R / W 0
loi_main_display_upper
Selects the measurement displayed on the upper por-
tion of the main display.
ENUM Assign PV
Blank
R / W Assign PV
loi_main_display_center
Selects the measurement displayed on the center por-
tion of the main display.
ENUM Compensating pH (free Cl)
Temperature
PV
Blank
R / W Temperature
loi_main_display_left
Selects the measurement or output displayed on the
left portion of the display.
ENUM Valid List per 1066 Model R / W
loi_main_display_lower_left
Selects the measurement or output displayed on the
lower left portion of the display.
ENUM Valid List per 1066 Model R / W
loi_main_display_right
Selects the measurement or output displayed on the
right portion of the display.
ENUM Valid List per 1066 Model R / W
loi_main_display_lower_right
Selects the measurement or output displayed on the
lower right portion of the display.
ENUM Valid List per 1066 Model R / W
1066pH/ORP Valid List
Input mV
Glass Impedance
Ref Impedance
Slope
Zero Offset
1066C/T Valid List
Raw Conductivity
Raw Resistivity
1066DO/CL/OZ Valid List
Input Current
pH (if free chlorine)
All 1066 Valid List
AO 1 Current
AO 1 % Output
AO 2 Current
AO 2 % Output