4 Specifications
Section 2: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-1066
Specifications - General
Case: Polycarbonate. IP66 (CSA, FM), Type 4X (CSA)
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x
139mm (5.45 x 5.45 in.)
Conduit openings: Six. Accepts PG13.5 or 1/2 in. conduit fittings
Display: Monochromatic graphic liquid crystal display. No backlight. 128 x 96 pixel display resolu-
tion. Active display area: 58 x 78mm (2.3 x 3.0 in.). All fields of the main instrument display can be
customized to meet user requirements.
Ambient temperature and humidity: -20 to 65 °C (-4 to 149°F), RH 5 to 95% (non-condensing).
Storage Temperature: -20 to 70 °F (-4 to 158 °F)
HART
®
Communications: PV, SV, TV, and 4V assignable to measurement, temperature and all live
HART diagnostics.
RFI/EMI: EN-61326-1
Complies with the following Standards:
CSA: C22.2 No 0 10; C22.2 No 0.4 04; C22.2 No. 25-M1966: , C22.2 No. 94-M91: , C22.2
No.142-M1987: , C22.2 No. 157-M1992: , C22.2 No. 213-M1987: , C22.2 No. 60529:05.
UL: 50:11th Ed.; 508:17th Ed.; 913:7th Ed.; 1203:4th Ed.. ANSI/ISA: 12.12.10-2013.
ATEX: EN 60079-0:2012+A11:2013, 60079-11:2012
IECEx: IEC 60079-0: 2011 Edition: 6.0, I EC 60079-11 : 2011-06 Edition: 6.0
FM: 3600: 2011, 3610: 2010, 3611: 2004, 3810: 2005, IEC 60529:2004, ANSI/ISA 60079-0: 2009,
ANSI/ISA 60079-11: 2009
2.2
Hazardous Location Approvals
Intrinsic Safety (with appropriate safety barrier):
Class I, II, III, Div. 1*
Groups A-G
T4 Tamb = -20 °C to 65 °C
Enclosure 4X, IP66
For Intrincically Safe Installation,
see drawing 1400669
1180 II 1 G
Baseefa11ATEX0195X
Ex ia IIC T4 Ga
T4 Tamb = -20 °C to 65 °C
Non-Incendive:
Class I, Div. 2, Groups A-D*
Dust Ignition Proof Class II & III, Div 1, Groups EFG
Class II & III, Div. 1, Groups E-G
Type 4/4X Enclosure
T4 Tamb = -20 °C to 65 °C
For Non-Incendive Field Wiring Installation, see drawing 1400669
ATEX
IECEx BAS 11.0098X
Ex ia IIC T4 Ga
T4 Tamb = -20 °C to 65 °C
Class I, II & III, Division 1, Groups A-G T4
Tamb = -20 °C to 65 °C
IP66 enclosure
Class I, Zone 0, AEx ia IIC T4
Tamb = -20°C to 65°C
For Intrinsically Safe Installation, see drawing 1400670
Class I, Division 2 Groups A-D
Dust Ignition proof Class II & III, Div 1, Groups EFG
Class II & III, Division 1, Groups E-G
Tamb = -20°C to 65°C, IP66 enclosure
For Non-Incendive Field Wiring Installation, see drawing 1400670
*Additionally approved as a system with models 140,141,142, 150, 400, 400VP, 401, 402, 402VP, 403,403VP, 404 & 410VP contacting conductivity
sensors and models 222, 225, 226 & 228 inductive conductivity sensors.