Certificate
IECEx BAS12.0073X
Standards
IEC60079-0:2011, IEC60079-31:2008
Markings
Ex t IIIC T50 °C T
500
60 °C Da
Table 9-5: Input Parameters
HART
Voltage U
i
36 V
Current I
i
24 mA
Special Conditions for Safe Use(X)
1. Cable entries must be used which maintain the ingress protection of
the enclosure to at least IP66.
2. Unused cable entries must be filled with suitable blanking plugs
which maintain the ingress protection of the enclosure to at least
IP66.
3. Cable entries and blanking plugs must be suitable for the ambient
temperature range of the apparatus and capable of withstanding a 7J
impact test.
9.5 Brazil
E2 INMETRO Flameproof
Certificate
UL-BR 15.0728X
Standards
ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-1:2016,
ABNT NBR IEC 60079-26:2016
Markings
Ex db IIC T6…T4 Ga/Gb T4/T5(–60 °C ≤ T
a
≤ +80 °C), T6(–60 °C
≤ T
a
≤ +70 °C)
Special Conditions for Safe Use(X)
1. This device contains a thin wall diaphragm less than 1 mm thickness
that forms a boundary between zone 0 (process connection) and
zone 1 (all other parts of the equipment). The model code and
datasheet are to be consulted for details of the diaphragm material.
Installation, maintenance, and use shall take into account the
environmental conditions to which the diaphragm will be subjected.
The manufacturer's instructions for installation and maintenance
shall be followed in detail to assure safety during its expected
lifetime.
2. Flameproof joints are not intended for repair.
Quick Start Guide May 2019
28 Emerson.com/Rosemount