EasyManua.ls Logo

Emerson Rosemount 3051SAL - Page 138

Emerson Rosemount 3051SAL
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
130
Reference Manual
00809-0200-4802, Rev BA
Appendix A: Specifications and reference data
May 2013
Specifications and Reference Data
Table A-11. Rosemount 3051S Liquid Level Transmitter
The Standard offering represents the most common models and options. These options should be selected for best delivery.
__The Expanded offering is subject to additional delivery lead time.
Model Transmitter Type
3051S Liquid Level Transmitter
Performance Class
Standard Standard
1 Ultra: 0.065% span accuracy, 100:1 rangedown, 12-year limited warranty
2 Classic: 0.065% span accuracy, 100:1 rangedown
Connection Type
Standard Standard
L Level
Measurement Type
Standard Standard
D Differential
G Gage
A Absolute
Pressure Range
Differential (LD) Gage (LG) Absolute (LA)
Standard Standard
2A -250 to 250 inH
2
O (-623 to
623 mbar)
-250 to 250 inH
2
O (-623 to
623 mbar)
0 to 150 psia (10 bar)
3A -1000 to 1000 inH
2
O (-2,5 to
2,5 bar)
-393 to 1000 inH
2
O (-0,98 to 2,5
bar)
0 to 800 psia (55 bar)
4A -300 to 300 psi (-20,7 to 20,7 bar) -14.2 to 300 psig (-0,98 to 21 bar) 0 to 4000 psia (276 bar)
5A -2000 to 2000 psi (-137,9 to
137,9 bar)
-14.2 to 2000 psig (-0,98 to 137,9
bar)
N/A
Transmitter Output
Standard Standard
A 4-20 mA with digital signal based on HART protocol
F
(1)
FOUNDATION fieldbus protocol
X
(2)
Wireless (Requires wireless options and wireless PlantWeb housing)
Housing Style Material Conduit Entry Size
Standard Standard
00 None (SuperModule spare part, order output code A)
1A PlantWeb housing Aluminum
1
/2–14 NPT
1B PlantWeb housing Aluminum M20 x 1.5
1J PlantWeb housing SST
1
/2–14 NPT
1K PlantWeb housing SST M20 x 1.5

Table of Contents