2.2  Troubleshooting the meter
Table 2-1 and the following sections show errors that may occur with the meter
hardware, firmware or connections and recommended actions to resolve the
problem(s).
Table 2-1: Troubleshooting
Error Recommended action(s)
Acquisition Module Error Check interconnect cable between Acquisition Module
and the CPU Module.
Attempt the Program Download procedure to install
the firmware.
Cycle power to the meter.
Replace the Acquisition Module.
If the Acquisition Module cannot be
reprogrammed, collect a complete Archive log
and contact your local area Emerson Flow service
representative.
Acquisition Module is not
compatible with firmware
Upgrade the firmware in the meter to the latest
version using MeterLink. Contact your Emerson Flow
service representative to obtain the latest firmware or
download the firmware from: www.emerson.com.
Replace the Acquisition Module.