Figure 3-19: Base Enclosure security seals
A. Transmitter Electronics Enclosure
B. Security wire seals
C. Transmitter Electronics endcap security latch
D. Base Enclosure
33. If required by the site operations manager, have an electrician fully test the
connections. After the Acceptance Test is witnessed and approved, seal the
conduit.
34. Power down the system and apply the sealing compound to the conduit
and allow to set in accordance with manufacturer specifications.
This completes the transducer cable removal and installation procedure.
3.5  Replace the meter electronics
The following procedure should be performed by a qualified service technician or
trained personnel. Observe all warning labels on the meter before starting this
procedure.
The Rosemount
3810 Series Ultrasonic Flow Meter Transmitter Electronics
Enclosure consists of the following:
CPU Module assembly (P/N 1-360-03-014)
Optional I/O Module (RS-232 or RS-485)
I.S. Barrier Board (P/N 1-360-03-004)
Maintenance and Troubleshooting Manual Meter maintenance
00809-0100-3814 January2023
Rosemount 3814 Liquid Ultrasonic Flow Meter 61