EasyManua.ls Logo

Emerson Rosemount 5400 Series - Page 23

Emerson Rosemount 5400 Series
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quick Start Guide
23
June 2016
2. WARNING – The apparatus enclosure contains aluminum and is considered to
constitute a potential risk of ignition by impact or friction. Care must be taken into
account during installation and use to prevent impact or friction.
7.6 Canada
E6 Explosionproof, Dust-Ignitionproof
Certificate: 1514653
Standards: CSA C22.2 No.0-M91, CSA C22.2 No.25-1966, CSA C22.2 No.30-M1986,
CSA C22.2 No.94-M91, CSA C22.2 No.142-M1987, CSA C22.2 157-92,
CAN/CSA C22.2 No. 60529:05, ANSI/ISA 12.27.01-2003
Markings: Explosionproof CL I, DIV 1, GP B, C, D; Dust-Ignitionproof CL II, DIV 1 and 2,
GP E, F, G and coal dust, CL III, DIV 1, Type 4X/IP66/IP67
I6 Intrinsically Safe and Non-Incendive Systems
Certificate: 1514653
Standards: CSA C22.2 No.0-M91, CSA C22.2 No.25-1966, CSA C22.2 No.30-M1986,
CSA C22.2 No.94-M91, CSA C22.2 No.142-M1987, CSA C22.2 157-92,
CAN/CSA C22.2 No. 60529:05, ANSI/ISA 12.27.01-2003
Markings: CL I, DIV 1, GP A, B, C, D, T4 see installation drawing 9150079-906;
Non-Incendive Class III, DIV 1, Haz-loc CL I DIV 2, GP A, B, C, D, Maximum
Ambient Temperature +60°C for Fieldbus and FISCO and +70 °C for HART,
T4, Type 4X/IP66/IP67, Maximum Working Pressure 5000 psi, Dual Seal.
IF FISCO
Certificate: 1514653
Standards: CSA C22.2 No.0-M91, CSA C22.2 No.25-1966, CSA C22.2 No.30-M1986,
CSA C22.2 No.94-M91, CSA C22.2 No.142-M1987, CSA C22.2 157-92,
CAN/CSA C22.2 No. 60529:05, ANSI/ISA 12.27.01-2003
Markings: CL I, DIV 1, GP A, B, C, D, T4 see installation drawing 9150079-906;
Non-Incendive Class III, DIV 1, Haz-loc CL I DIV 2, GP A, B, C, D, Maximum
Ambient Temperature +60°C for Fieldbus and FISCO and +70 °C for HART,
T4, Type 4X/IP66/IP67, Maximum Working Pressure 5000 psi, Dual Seal.
7.7 Europe
E1 ATEX Flameproof
Certificate: Nemko 04ATEX1073X
Standards: EN 60079-0:2012, EN 60079-1:2014, EN 60079-11:2012,
EN 60079-26:2015, EN 60079-31:2014
Markings: II 1/2 G Ex db ia IIC T4 Ga/Gb, (-40°C Ta +60°C /+70°C)
II 1 D Ex ta IIIC T69°C/T79°C Da, (-40°C Ta +60°C /+70°C)
Um = 250 V
Ui Ii Pi Ci Li
FISCO parameters
17.5 V 380 mA 5.32 W 0 0
Ui Ii Pi Ci Li
Entity parameters HART
30 V 130 mA 1 W 7.26 nF 0
Entity parameters Fieldbus
30 V 300 mA 1.3 W 0 0
Ui Ii Pi Ci Li
FISCO parameters
17.5 V 380 mA 5.32 W 0 0