EasyManua.ls Logo

Emerson Rosemount 8600D Series - Calculating Output Variables with Known; Input Frequency123

Emerson Rosemount 8600D Series
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123
Reference Manual
00809-0100-4860, Rev BC
Appendix C: Electronics verification
January 2013
Electronics verification
Figure C-2. Test Frequency Output and Chassis Ground Points.
C.2.5 Calculating output variables with known
input frequency
Use the following equations with a known input frequency for verification of a flow rate or 4–20 mA
output within a given calibrated range. Select the proper equation depending on if you are verifying
a flow rate, mass flow rate, 420 mA output, or special units. Example calculations starting on
page 125 may clarify how these equations are used.
To verify a flow rate
For a given frequency F (Hz), and K-factor (compensated), find the
flow rate Q:
where C
x
is the unit conversion (Table C-1 on page C-125).
To verify a standard or normal flow rate
To verify a mass flow rate
For a given mass frequency F (Hz), and K-factor (compensated), find the mass flow rate M:
where C is the unit conversion and
is density at operating conditions:
where C
x
is the unit conversion using density () (Table C-1 on page C-125).
Ground
Test Frequency In
Test Point 1
QFHzKC
x
=
QFHz DensityRatio KC
x
 =
M
F
K  C
-------------------------=
MFHzKC
x
=
Get other manuals https://www.bkmanuals.com