EasyManua.ls Logo

Emerson Rosemount 8600D Series - SI Units

Emerson Rosemount 8600D Series
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
128
Reference Manual
00809-0100-4860, Rev BC
Appendix C: Electronics verification
January 2013
Electronics verification
C.3.2 SI units
Example 1 (SI units)
Therefore, an input frequency of 80.00 Hz represents a flow rate of 1694.6 lpm in this
application.
For a given input frequency, you may also determine the current output. Use the above example
with an input frequency of 80.00 Hz:
URV= 2000 lpm LRV= 0 lpm F
in
(Hz) = 80.00
= 17.49 mA
Therefore, an input frequency of 80.00 Hz represents a current output
of 17.49 mA.
Fluid = Water URV = 2000 lpm
Line size = 80 mm. LRV = 0 lpm
Line pressure = 700 kPas C
8
= 4.40287E-03 (from
Table C-1 on
page C-125
)
Operating
Tem peratu re
= 60 °F
Input
Frequency
= 80 Hz
K-factor
(compensated)
= 10.772 (via Field Communicator or AMS)
M = F(Hz) /(K x C
8
)
= 80/(10.722 x 0.00440287)
= 1694.6 lpm
I
FHzKC
8
 LRV
URV LRV
-------------------------------------------------------------
16
4+=
I
80 10.772 0.004402870
2000 0
--------------------------------------------------------------------------------------
16



4+=