Table 2-2: Backflush Valve Timing
PA Old
RT BF TIME PA n-Pentane % Change
219 42 76822312
219 42.5 82954542 7.982355439
219 43 85914053 3.567629847
219 43.5 88965346 3.551564492
219 44 92305383 3.754312381
219 44.5 93208291 0.978174805
219 45 94027248 0.878631065
219 45.5 93370502 0.698463492
219 46 94646236 1.36631374
Retrofit Instructions Installation
SK-07781 September 2019
Rosemount Model 500 Helium to Hydrogen Gas Conversion Kit 19