EMS AIR-FLOW MASTER Perio - Troubleshooting

EMS AIR-FLOW MASTER Perio
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80
Symptom Solution
The unit does not start or stop • Check the electrical connections and power supply at the mains
• Check the fuses on back panel
• Check main switch function: at least one LED must be visible on adjustments and working mode areas
• The unit is in "standby" mode, tap on any sensitive zone to exit standby
• Send the unit to an authorized EMS Repair Center
The foot pedal does not activate the unit • Check that the foot pedal is properly connected
• Replace the foot pedal
• Send the unit to an authorized EMS Repair Center
The unit is still activated when the foot
pedal is released
• The foot pedal is active, check for jam at foot pedal
• The switch in the pedal is blocked. Turn the unit off at the power switch
• Replace the foot pedal
• Send the unit to an authorized EMS Repair Center
User interface displays an error: there
is one acoustic signal and all LEDs are
blinking
• Powder chamber pressure is too low or too high
• The air or liquid regulator is switched off
• Foot pedal and purge commands are not recognized by system. Check connections
The handpiece does not remain on the
magnetic support
• Install the magnetic support on a metallic part of the unit
• The support is demagnetized
• Replace magnetic support
The handpiece does not hold on the
handpiece cord
• Check for presence of the O-ring seal and replace if necessary
• Replace the handpiece
• Replace the handpiece cord
• Send the handpiece and the handpiece cord to an authorized EMS Repair Center
No water jet coming from the handpiece • Check the water supply and water lter
• Check the water adjustment of your unit
• Check the connections between the unit, the cord and the handpiece
• Check for presence of water at the cord outlet port
• Send the handpiece to an authorized EMS Repair Center
No powder/air jet coming from the
handpiece
• Check the quantity of powder in your unit
• Check the compressed air supply and air lter
• Check for presence of powder/air at cord outlet port
• Eliminate residual powder present in the channels using the cleaning needles provided
• Perio powder chamber only: remove the diffuser and empty the powder. Blow the diffuser with compressed air before
installing it again and relling.
• Send the system to an authorized EMS Repair Center
TROUBLESHOOTING
Cont. next page
FB-580_CN_rev_B_01_Not Util AIR-FLOW MASTER.indd 80 13.10.2017 10:00:48