Cleaning the handpiece with the Easy Clean
The EASY CLEAN is designed to unclog and to facilitate the cleaning of the EMS AIR-FLOW handpieces.
The EASY CLEAN can be thermal disinfected and also sterilized at up to 135°C in the autoclave.
Periodical cleaning of the handpiece using the EASY CLEAN.
Rinse through central lumina in the normal direction of ow (no back rinsing) using a disposable syringe lled with
more than 2 ml of drinking water quality applied to the purge adaptor.
After removing the EASY CLEAN, blow compressed air into the handpiece until completely dry inside the handpiece.
It is recommended to perform a periodical cleaning once a week o after 15 treatments.