40
Symptom Solutions
No water jet coming from the
handpiece
Check the water adjustment of your unit
Check the connections between the unit, the cord and the handpiece
Check the water supply
Check for presence of water at the cord outlet port
Check for presence of water at the system connector outlet without cord
Send the system to an approved EMS repair center
No powder/air spray coming from the
handpiece
Check the compressed air supply
Check for presence of powder/air at cord outlet port without handpiece
Check for presence of powder/air at the system connector outlet without cord
Eliminate residual powder present in the channels using cleaning needles or the Easy Clean
Check the quantity of powder in your unit
Send the system to an approved EMS repair center
Water leakage between handpiece cord
and handpiece
Check the condition of the handpiece connecting O-rings
Send the system and the handpiece cord to an approved EMS repair center
The handpiece efciency decreases Check the quantity of powder in the unit
Eliminate residual powder present in the channels using cleaning needles or the Easy Clean
Send the system and the handpiece cord to an approved EMS repair center
TROUBLESHOOTING