EMS AIR-FLOW - Troubleshooting

EMS AIR-FLOW
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
Symptom Solutions
No water jet coming from the
handpiece
• Check the water adjustment of your unit
• Check the connections between the unit, the cord and the handpiece
• Check the water supply
• Check for presence of water at the cord outlet port
• Check for presence of water at the system connector outlet without cord
• Send the system to an approved EMS repair center
No powder/air spray coming from the
handpiece
• Check the compressed air supply
• Check for presence of powder/air at cord outlet port without handpiece
• Check for presence of powder/air at the system connector outlet without cord
• Eliminate residual powder present in the channels using cleaning needles or the Easy Clean
• Check the quantity of powder in your unit
• Send the system to an approved EMS repair center
Water leakage between handpiece cord
and handpiece
• Check the condition of the handpiece connecting O-rings
• Send the system and the handpiece cord to an approved EMS repair center
The handpiece efciency decreases • Check the quantity of powder in the unit
• Eliminate residual powder present in the channels using cleaning needles or the Easy Clean
• Send the system and the handpiece cord to an approved EMS repair center
TROUBLESHOOTING