20
Solutions compatible with the circuit
Incompatible mixing
Different solutions can be used during treatment. Chemical reactions might occur between these
solutions and damage the unit. Wash the liquid circuit for 20 seconds with distilled (or demineralized) water
before using a different solution (please refer to the "End of treatment" chapter).
Irigation solutions for the treatment
• Chlorhexidine (≤ 0.3%)
• Physiological water (NaCl ≤ 0.9%)
• Hydrogen peroxide (H
2
O
2
≤ 3%)
• Carbanilides
• Quaternary ammonium compounds
• Citric acid
• Aromatic essence
• Ethanol
Examples of common products
BacterX
®
pro / Triclosan / Listerine
®
/ Alodont
®
/ Meridol
®
/ BreathRx
®
/ CloSYS™ / Scope
®
/ Cepacol
®
/.....
Solutions for washing the circuit
• EMS Ultra Clean (H
2
O
2
6%)
• Hydrogen peroxide (H
2
O
2
5%)
• Sterilox (hypochlorous acid)
Please read the instructions for use of the solutions.
Please refer to standards applicable in your country regarding the purchase and use of these
products.