70
TROUBLESHOOTING
Symptom Solution
The LEDs of the power supply do not
illuminate when the unit is switched
on
Check that the mains cord of the power supply is properly connected to the mains supply
Check the power supply of the dental ofce
Turn the power knob of the unit to ON and check if the LED is lit
Replace the power supply
Send the unit and the accessories to an approved EMS repair center
No irrigation ow when the foot
switch is pressed
Turn the liquid ow rate knob
Make sure that water comes in external device for the PM100
Make sure that the bottle is properly installed for the PM200
Disconnect the ultrasound handpiece and check that the liquid exits from the cord
Use compressed air (max. 3 bar) and check that air comes out of the handpiece
If the air does not come out of the handpiece send to an approved EMS repair center
Check the connection of the foot switch
Replace the foot switch
Replace the lter for the PM100
Replace the pump for the PM200
Send the unit and the accessories to an approved EMS repair center
No ultrasound oscillations when the
foot switch is pressed
Check that the handpiece is connected to the handpiece cord
Control that the instrument is tight
Check that the instrument is not cracked
Check the connection of the foot switch
Replace the foot switch
Send the unit and the accessories to an approved EMS repair center
Ultrasound power decreasing or
inadequate
Check that the selector is beyond the 4-position
Control that the instrument is tight
Check the wear of the instrument and replace if necessary
Replace the handpiece
Send the unit and the accessories to an approved EMS repair center
Water leaks between the handpiece
cord and the handpiece
Replace the O-ring on the handpiece cord
Replace the handpiece
Send the handpiece to an approved EMS repair center
Only PM200:
Water leaks from the bottle or the
device
Check the O-ring of the bottle cap
Replace the bottle
Send the handpiece to an approved EMS repair center
FB-544-5_Notice_PM100_PM200.indd 70 3/27/2013 10:46:36 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS PM100 and is the answer not in the manual?

EMS PM100 Specifications

General IconGeneral
BrandEMS
ModelPM100
CategoryDental equipment
LanguageEnglish

Related product manuals