Installation and Setup TOPwave
22
upstream wash bottle). In addition to this the oven chamber is permanently sucked off
and the gases must be conveyed to the extraction system via the PE corrugated tube
supplied.
1 Connection PE exhaust tube
(ø 62 mm)
2 Connection PFA exhaust tube
(ø 10 mm)
WARNING
Hot acid vapors may be released.
The supplied vent tubes must therefore alwa
ys be properly connected and attached to
sure the PFA tube is securely attached so that it cannot pop off its connection fit-
ting in case of a percussive pressure burst as occurs when a rupture disc activates
.
NOTICE
Provide a suitable vent!
Users are responsible for assuring that appropriate v
entilation of toxic gases is assured.
Make sure that all ventilation procedures conform to applicable laws.
Use the supplied RS-232 cable to connect the basic unit with separate control unit.
Connect both units with supplied power cord to a suitable 230 V; 50/60 Hz power
source.
WARNING
Protective conductor terminal
.
Basic unit and separate control unit must be connected to the same phase of mains and
always be correct connected to protective earth.
The power switch is located on the right hand side of the unit.
The switch will be illuminated when it is on. The fuse and the RS-232 connection are ar-
ranged next to the switch. The RS-232 connection is the PC-interface.
Connection
Power switch (microwave)